TIM Tester Calibration Package:
Calibration Instruments & Reference Sample Set
TIM Tester instrument calibrations can be conveniently performed "in-lab" without ever moving the unit. The TIM Calibrator and TC Calibrator accessories are provided for this purpose. The entire bi-annual calibration of the TIM Tester is a user-friendly, 10 minute procedure handled by WinTIM Software. During calibration, the TIM Calibrator is connected to the rear panel of the TIM Tester and provides calibration signals which can be measured with a calibrated lab-voltmeter (not included). The TC Calibrator provides calibration of the TIM thermocouple ports in conjunction with WinTIM cal-routines. WinTIM also provides complete help files and on-screen instructions for all aspects of instrument calibration. The Tc Calibrator and TIM Calibrator should be sent for yearly calibration and re-certification to Analysis Tech.
The Reference Sample Set provides a collection of prepared, pre-tested reference samples spanning the appropriate range of thermal impedances. These reference samples can be tested at any time to confirm the accuracy and consistency of the TIM Tester operation. Reference samples are tagged with the sample's impedance, tolerance, and specified test pressure. The included documentation provides certified traceability to national laboratory standards. Reference standard samples provide an excellent and convenient means to maintain highest verification and confidence in the TIM Test results.
This package is included with the TIM Tester, and also includes the Complete Plumbing Kit and Sample Positioning Fixture.